# 8/5/2021 Only sensor 2 is working. Running with dspIC # 20/9/2021 sensors 1,2,3 are working. Running with dspIC # 20/9/2021 sensors 1,2,3 are working. Running with tty USB record (longin, "$(P)$(R)P001:M") { field (DESC, "getPressure001 $(bus)") field (DTYP, "stream") field (INP, "@pfeiffer.proto getPressure001($(P)$(R)P001:E) $(bus)") # field (SCAN, "I/O Intr") field (SCAN, "1 second") field (FLNK, "$(P)$(R)TMPump1-PressureAdmission") } record (longin, "$(P)$(R)P001:E") { field (DESC, "Pressure001 exponent $(bus)") } record (calc, "$(P)$(R)TMPump1-PressureAdmission") { field(DESC, "ISTTOK:central:TMPump1-PressureAdmission") field(INPA, "$(P)$(R)P001:M") field(INPB, "$(P)$(R)P001:E") field(CALC, "(A*0.001)* 10.0**(B-20)") field(PREC, "3") field(EGU, "mBar") field(HIHI,"1.0e-2") field(HIGH,"1.0e-3") # field(LOW, "20.0") # field(LOLO,"0.0") field(HHSV,"MAJOR") field(HSV, "MINOR") # field(LSV, "MINOR") # field(LLSV,"MAJOR") } # out "0020074002=?107"; record (longin, "$(P)$(R)P002:M") { field (DESC, "getPressure002 $(bus)") field (DTYP, "stream") field (INP, "@pfeiffer.proto getPressure002($(P)$(R)P002:E) $(bus)") # field (SCAN, "I/O Intr") field (SCAN, "1 second") field (FLNK, "$(P)$(R)RPump1-Pressure") } record (longin, "$(P)$(R)P002:E") { field (DESC, "Pressure002 exponent $(bus)") } record (calc, "$(P)$(R)RPump1-Pressure") { field (DESC, "ISTTOK:central:RPump1-Pressure") field(INPA, "$(P)$(R)P002:M") field(INPB, "$(P)$(R)P002:E") field(CALC, "(A*0.001)*10.0**(B-20)") field(PREC, "3") field(EGU, "mBar") field(HIHI,"1.0e-1") field(HIGH,"5.0e-2") # field(LOW, "20.0") # field(LOLO,"0.0") field(HHSV,"MAJOR") field(HSV, "MINOR") # field(LSV, "MINOR") # field(LLSV,"MAJOR") } # out "0030074002=?108"; record (longin, "$(P)$(R)P003:M") { field (DESC, "getPressure003 Mantissa $(bus)") field (DTYP, "stream") field (INP, "@pfeiffer.proto getPressure003($(P)$(R)P003:E) $(bus)") field (SCAN, "1 second") # field (SCAN, "I/O Intr") field (FLNK, "$(P)$(R)VVessel-Pressure") } record (longin, "$(P)$(R)P003:E") { field (DESC, "Pressure003 Exponent $(bus)") } record (calc, "$(P)$(R)VVessel-Pressure") { field (DESC, "ISTTOK:central:VVessel-Pressure") field(INPA, "$(P)$(R)P003:M") field(INPB, "$(P)$(R)P003:E") field(CALC, "(A*0.001)*10.0**(B-20)") field(PREC, "3") field(EGU, "mBar") field(HIHI,"1.0e-1") field(HIGH,"1.0e-2") field(HHSV,"MAJOR") field(HSV, "MINOR") } # out "0040074002=?109"; # NOT connected 26/10/2021 record (longin, "$(P)$(R)P004:M") { field (DESC, "getPressure004 Mantissa $(bus)") field (DTYP, "stream") field (INP, "@pfeiffer.proto getPressure004($(P)$(R)P004:E) $(bus)") # field (SCAN, "1 second") # field (SCAN, "I/O Intr") field (FLNK, "$(P)$(R)RPump2-Pressure") } record (longin, "$(P)$(R)P004:E") { field (DESC, "Pressure004 Exponent $(bus)") } record (calc, "$(P)$(R)RPump2-Pressure") { field(DESC, "ISTTOK:central:RPump2-Pressure") field(INPA, "$(P)$(R)P004:M") field(INPB, "$(P)$(R)P004:E") field(CALC, "(A*0.001)* 10.0**(B-20)") field(PREC, "3") field(EGU, "mBar") # field(HIHI,"1.0e-1") # field(HIGH,"5.0e-3") # field(LOW, "20.0") # field(LOLO,"0.0") # field(HHSV,"MAJOR") # field(HSV, "MINOR") # field(LSV, "MINOR") # field(LLSV,"MAJOR") } record(stringout,"$(P)$(R)RS485Debug") { field(DESC, "") field(DTYP, "stream") field(OUT, "@pfeiffer.proto debug $(bus)") field(VAL,"") }